翻訳と辞書
Words near each other
・ "O" Is for Outlaw
・ "O"-Jung.Ban.Hap.
・ "Ode-to-Napoleon" hexachord
・ "Oh Yeah!" Live
・ "Our Contemporary" regional art exhibition (Leningrad, 1975)
・ "P" Is for Peril
・ "Pimpernel" Smith
・ "Polish death camp" controversy
・ "Pro knigi" ("About books")
・ "Prosopa" Greek Television Awards
・ "Pussy Cats" Starring the Walkmen
・ "Q" Is for Quarry
・ "R" Is for Ricochet
・ "R" The King (2016 film)
・ "Rags" Ragland
・ ! (album)
・ ! (disambiguation)
・ !!
・ !!!
・ !!! (album)
・ !!Destroy-Oh-Boy!!
・ !Action Pact!
・ !Arriba! La Pachanga
・ !Hero
・ !Hero (album)
・ !Kung language
・ !Oka Tokat
・ !PAUS3
・ !T.O.O.H.!
・ !Women Art Revolution


Dictionary Lists
翻訳と辞書 辞書検索 [ 開発暫定版 ]
スポンサード リンク

Therac 25 : ウィキペディア英語版
Therac-25
The Therac-25 was a radiation therapy machine produced by Atomic Energy of Canada Limited (AECL) after the Therac-6 and Therac-20 units (the earlier units had been produced in partnership with CGR of France).
It was involved in at least six accidents between 1985 and 1987, in which patients were given massive overdoses of radiation. Because of concurrent programming errors, it sometimes gave its patients radiation doses that were thousands of times greater than normal, resulting in death or serious injury. These accidents highlighted the dangers of software control of safety-critical systems, and they have become a standard case study in health informatics and software engineering.
== Design ==

The machine offered two modes of radiation therapy:
* Direct electron-beam therapy, which delivered low doses of high-energy (5 MeV to 25 MeV) electrons over short periods of time;
* Megavolt X-ray therapy, which delivered X-rays produced by colliding high-energy (25 MeV) electrons into a "target".
When operating in direct electron-beam therapy mode, a low-powered electron beam was emitted directly from the machine, then spread to safe concentration using scanning magnets. When operating in megavolt X-ray mode, the machine was designed to rotate four components into the path of the electron beam: a target, which converted the electron beam into X-rays; a flattening filter, which spread the beam out over a larger area; a set of movable blocks (also called a collimator), which shaped the X-ray beam; and an X-ray ion chamber, which measured the strength of the beam.

抄文引用元・出典: フリー百科事典『 ウィキペディア(Wikipedia)
ウィキペディアで「Therac-25」の詳細全文を読む



スポンサード リンク
翻訳と辞書 : 翻訳のためのインターネットリソース

Copyright(C) kotoba.ne.jp 1997-2016. All Rights Reserved.